Abbott Medical recalls
7 recalls on record · 4 rated Critical · FDA
Every recall we have on record for Abbott Medical, compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
Abbott Infinity 7, Implantable Pulse Generator, REF: 6663, SterileEO, Rx Only
Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.
Abbott Proclaim XR 5 Implantable Pulse Generator, REF 3660, Sterile EO, Rx Only
The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.
Abbott Proclaim 5 Implantable Pulse Generator, REF 3661, Sterile EO, Rx Only
The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.
Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The…
Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).
Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5…
Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).
Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable…
Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).
Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable…
Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).