Aesculap Implant Systems LLC recalls
4 recalls on record · FDA
Every recall we have on record for Aesculap Implant Systems LLC, compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
Filter:
Health Serious
Aesculap ENNOVATE SET SCREW STERILE-implants are used for dorsal monosegmental and…
Potential for the incorrect screw to be included in the package.
Aesculap Implant Systems LLC Sep 7, 2022
Read the recall
Health Serious
PRESTIGE RETRACTION GRASPER, W/PORT DUAL ACTION, ROTATING 5mm DIAM., 36cm LENGTH. Product…
Endoscopic graspers may become separated at the weld from handle to the shaft of the device.
Aesculap Implant Systems LLC Aug 10, 2022
Read the recall
Health Serious
AESCULAP S4 Cervical Cross Connector L24mm, S4C CROSS CONNECTOR FIXED 24MM, product code…
Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion.
Aesculap Implant Systems LLC Jul 27, 2022
Read the recall
Health Serious
ENDURO SPECIAL KEY TIB.LOCK.RING F1/10MM product code NP462R used with EnduRo Total Knee…
Nonfunctional key due to the potential for damage to the key pins cause by force of use during assembly of the tibia locking ring.
Aesculap Implant Systems LLC Jul 27, 2022
Read the recall