Canon Medical System, USA, INC. recalls
2 recalls on record · FDA
Every recall we have on record for Canon Medical System, USA, INC., compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
System: Alphenix 4D CT in combination with CAS-930A (C arm Support) Model Number: …
CT operation may be restricted by an interlock which is a result of an system error and the CT system may stop operating properly after an attempt to cancel the error message has been initiated.