Cardinal Health 200, LLC recalls
12 recalls on record · 4 rated Critical · FDA
Every recall we have on record for Cardinal Health 200, LLC, compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
Salem Sump PVC Tubes: Product Code/Product Name: 8888265108 Salem Sump Silicone Dual…
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
Cardinal Health Presource Kits: 1) KIT,NASAL SEPTOPLASTY,FAIRVIEW, Catalog…
Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.
Cardinal Health Presource Kits: 1) KIT,DIEP FLAP , Catalog…
Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.
(a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch…
The firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.
Presource kits and trays, labeled as: 1) Cardiovascular Procedure Kit, Catalog Number…
Cardinal Health is following the FDA's recommendations regarding the disposition of affected Jiangsu Shenli Medical Production Co. Ltd (Shenli) Syringes in the Merit Medical¿ supplied finished kits. Merit Medical¿ supplied finished kits are included as components in Presource kits. According to the FDA s communication, these Shenli syringes, located inside a Merit Medical¿ kit component, cannot be distributed for use as they could contain a significant change or modification in design which could significantly alter the safety or effectiveness of the finished kit product, resulting in the risk of patient harm such as inaccurate dosing and a leaking device.
Jackson-Pratt 3-Spring Reservoir Kit with PVC Round Drain, 400mL, 7Fr., REF SU130-401D
Cardinal Health is issuing a lot number specific voluntary removal of Jackson-Pratt 3-Spring Reservoir kits due to specific products being shipped to users before undergoing sterilization.
Dover" IRRIGATION 1200 CC TRAY, 60 CC PISTON SYRINGE, STERILE SALINE, TIP PROTECTOR,…
Convenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
TRACH KIT W/HYDROGEN & SALI
Convenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
Dover" 100% SILICONE 400 ML U/M FOLEY TRAY, NEEDLE PORT, 16 FR (5.3 MM), 5 CC, 2000 ML…
Convenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
CardinalHealth Suction Canister Liner, MEDI-VAC CRD, REF 65651-510
The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.
Monoject 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack
Demonstrated recognition and compatibility issues with syringe infusion pumps.
Procedure pack: Presource PBDS Cat. PLMHCSBHF, Hermann Hospital MHHS, Kit, C Section B…
Packs contain recalled components (equipment drapes, table cover, and surgical room turnover (SRT) kits).