GE Medical Systems China Co., Ltd. recalls
3 recalls on record · 2 rated Critical · FDA
Every recall we have on record for GE Medical Systems China Co., Ltd., compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic…
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
GE Healthcare CARESTATION 620 A2, Model/REF Number 1012-9620-002. Anesthesia system.
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
GE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.