GE Vingmed Ultrasound As recalls
2 recalls on record · FDA
Every recall we have on record for GE Vingmed Ultrasound As, compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
Filter:
Health Serious
GE Vivid Ultrasound products, Model Numbers: a) Vivid S70, H45041SW
GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.
GE Vingmed Ultrasound As Jul 17, 2024
Read the recall
Health Serious
GE Healthcare Vivid S60N, ultrasound device
GE HealthCare has become aware that certain Vivid ultrasound systems cannot boot up in a timely fashion. If this occurs, it can delay availability of the device in time-critical emergency situations.
GE Vingmed Ultrasound As Jul 19, 2023
Read the recall