RecallCheckr

Ion Beam Applications S.A. recalls

2 recalls on record · FDA

Every recall we have on record for Ion Beam Applications S.A., compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.

Filter:
Health Serious

IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions:…

Safety Parameters Verification Mechanisms can be deactivated in clinical runtime

Ion Beam Applications S.A. Sep 18, 2024
Read the recall
Health Serious

Proteus235- Proton Therapy System: to produce and deliver a proton beam for the…

First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed

Ion Beam Applications S.A. Jun 8, 2022
Read the recall