Johnson & Johnson Surgical Vision, Inc. recalls
2 recalls on record · FDA
Every recall we have on record for Johnson & Johnson Surgical Vision, Inc., compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber …
Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.
VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)
Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.