Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls

21 recalls on record · 20 rated Critical · FDA

Every recall we have on record for Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.

Filter:
Health Critical

ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

CRT-D DTBB2D4 VIVA S IS1/DF4 INTL, Model Number DTBB2D4; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

CRT-D DTBA2QQ VIVA QUAD XT IS4/DF4 INTL, Model Number DTBA2QQ; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

CRT-D DTBA1Q1 VIVA QUAD XT IS4/DF1 US, Model Number DTBA1Q1; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

CRTD CROME HF MRI IS1 DF4, Model Number DTPC2D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

CRTD DTMA2D1 CLARIA MRI OUS DF1, Model Number DTMA2D1; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

CRTD DTMC1D1 COMPIA MRI US DF1, Model Number DTMC1D1; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

CRTD DTMB1D4 AMPLIA MRI US DF4, Model Number DTMB1D4; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

ICD-VR VISIA AF S US/OUS IS1/DF1, Model Number DVAC3D1; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

CRT-D DTBC2QQ BRAVA QUAD IS4/DF4 INTL, Model Number DTBC2QQ; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

ICD COBALT XT HF QUAD OUS, Model Number DTPA2QQG; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

ICD-DR DDBB2D4 EVERA XT IS1/DF4 INTL, Model Number DDBB2D4; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

ICD CROME VR MRI IS1 DF1, Model Number DVPC3D1; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

CRT-D DTBA1D1 VIVA XT IS1/DF1 US, Model Number DTBA1D1; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

CRTD DTMC2QQ COMPIA MRI QUAD OUS DF4, Model Number DTMC2QQ; Implantable Cardioverter…

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

ICD-VR DVME3D4 MIRRO MRI, Model Number DVME3D4; Implantable Cardioverter Defibrillators

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Jun 28, 2023
Read the recall
Health Critical

Crome Implantable Cardioverter Defibrillators: a. Crome VR ICD, Model Numbers: DVPC3D1,…

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Aug 17, 2022
Read the recall
Health Critical

Cobalt Implantable Cardioverter Defibrillators: a. Cobalt XT VR ICD, Model Numbers:…

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Aug 17, 2022
Read the recall
Health Serious

Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy…

Manufacturing error that may have resulted in a cathode component being out of specification. All devices met final functional testing requirements.

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) Aug 3, 2022
Read the recall