Health Serious

MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, #…

Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.

Micro-X Ltd. Dec 18, 2024
Read the recall
Health Serious

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Micro-X Ltd. Dec 18, 2024
Read the recall
Health Serious

MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions

Micro-X Ltd. Oct 5, 2022
Read the recall