Micro-X Ltd. recalls
3 recalls on record · FDA
Every recall we have on record for Micro-X Ltd., compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, #…
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
MICRO-X Rover Mobile X-ray System, # MXU-RV35
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19
Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions