RecallCheckr

RAYSEARCH LABORATORIES AB recalls

9 recalls on record · FDA

Every recall we have on record for RAYSEARCH LABORATORIES AB, compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.

Filter:
Health Serious

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0,…

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

RAYSEARCH LABORATORIES AB Jan 28, 2026
Read the recall
Health Serious

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0,…

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

RAYSEARCH LABORATORIES AB Jan 28, 2026
Read the recall
Health Serious

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0,…

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

RAYSEARCH LABORATORIES AB Jan 28, 2026
Read the recall
Health Serious

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 16.0.0 …

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

RAYSEARCH LABORATORIES AB Jan 28, 2026
Read the recall
Health Serious

RayStation with the following product descriptions: 1. RayStation 11B, Software…

DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

RAYSEARCH LABORATORIES AB Sep 10, 2025
Read the recall
Health Serious

RayStation with the following product descriptions: 1. RayStation 8B, Software…

DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

RAYSEARCH LABORATORIES AB Sep 10, 2025
Read the recall
Health Serious

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System …

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

RAYSEARCH LABORATORIES AB Feb 26, 2025
Read the recall
Health Serious

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System …

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

RAYSEARCH LABORATORIES AB Feb 26, 2025
Read the recall
Health Serious

RayStation 13.0.0.1547, 13.1.0.144 & 13.1.1.89. Radiation Therapy Treatment Planning…

Potential for reported SSD to be too high.

RAYSEARCH LABORATORIES AB May 8, 2024
Read the recall