Tandem Diabetes Care, Inc. recalls
3 recalls on record · 2 rated Critical · FDA
Every recall we have on record for Tandem Diabetes Care, Inc., compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)
an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.
Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; …
Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.
t:slim X2 Insulin Pump with Interoperable Technology
A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.