Wright Medical Technology, Inc. recalls
3 recalls on record · FDA
Every recall we have on record for Wright Medical Technology, Inc., compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
Stryker Infinity Alignment Frame Distal Sub-Assembly, Catalog #33600020, non-sterile.
The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.
ORTHOLOC 3DSi CLAW II 4 HOLE Plate, Size 25MM, REF 40241425, Bone fixation device
Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled, resulting in the incorrect product in packages. Packaging for 402S0425 incorrectly contained 40241425, and vice versa.
Wright EVOLVE TRIAD Plate Cutter, REF 49510120
The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use