Siemens Medical Solutions USA, Inc 2022-04-27 Health serious FDA
Health · FDA · Recall #FDA-Z-0971-2022

Artis zee floor MN, Model Number 10094142 Recall

Issued April 27, 2022 · Siemens Medical Solutions USA, Inc

In April 2022, Siemens Medical Solutions USA, Inc recalled the Artis zee floor MN, Model Number 10094142, a serious health recall reported by the FDA. Affected owners are eligible for a voluntary. Here is what the hazard is and exactly what to do about it.

Compiled by Maya Ellison, RecallCheckr

What this means for you

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

What you should do

  1. 1 Stop using Artis zee floor MN, Model Number 10094142 right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Siemens Medical Solutions USA, Inc to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Siemens Medical Solutions USA, Inc
Category
Health
Recall date
April 27, 2022
Units affected
Not reported
Sold at
US Nationwide.
Recall number
FDA-Z-0971-2022
Made in
United States
Issuing agency
FDA