RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Atrium Medical Corporation 2024-01-03 Health serious FDA
Health · FDA · Recall #FDA-Z-0621-2024

Atrium Oasis Drain Single W / AC, Part No. 3600-100 Recall

Issued January 3, 2024 · Atrium Medical Corporation

In January 2024, Atrium Medical Corporation recalled the Atrium Oasis Drain Single W / AC, Part No. 3600-100, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

A sterilization nonconformance prematurely aged the product by 48.3 days, resulting in the expiration date on the label not being accurate. This nonconforming product was inadvertently distributed to customers. Increased risk to patients and healthcare providers using the product after the shortened expiration date of November 17, 2025.

What you should do

  1. 1 Stop using Atrium Oasis Drain Single W / AC, Part No. 3600-100 right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Atrium Medical Corporation to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Atrium Medical Corporation
Category
Health
Recall date
January 3, 2024
Units affected
Not reported
Sold at
Domestic distribution nationwide. No foreign distribution.
Recall number
FDA-Z-0621-2024
Made in
United States
Issuing agency
FDA

In context

  • Atrium Medical Corporation has 9 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 737th Health recall we've logged in 2024.
Read the official FDA notice

Related recalls