RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

DePuy Orthopaedics, Inc. 2022-07-27 Health serious FDA
Health · FDA · Recall #FDA-Z-1419-2022

ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the… Recall

Issued July 27, 2022 · DePuy Orthopaedics, Inc.

In July 2022, DePuy Orthopaedics, Inc. recalled the ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the…, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain

What you should do

  1. 1 Stop using ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the… right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact DePuy Orthopaedics, Inc. to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
DePuy Orthopaedics, Inc.
Category
Health
Recall date
July 27, 2022
Units affected
Not reported
Sold at
Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finla
Recall number
FDA-Z-1419-2022
Made in
United States
Issuing agency
FDA

In context

  • DePuy Orthopaedics, Inc. has 4 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 870th Health recall we've logged in 2022.