RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Philips North America Llc 2026-02-18 Health serious FDA
Health · FDA · Recall #FDA-Z-1315-2026

Brilliance iCT; Product Code (REF): 728306; Recall

Issued February 18, 2026 · Philips North America Llc

In February 2026, Philips North America Llc recalled the Brilliance iCT; Product Code (REF): 728306;, a serious health recall reported by the FDA. Affected owners are eligible for a voluntary. Here is what the hazard is and exactly what to do about it.

Compiled by Maya Ellison, RecallCheckr

What this means for you

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

What you should do

  1. 1 Stop using Brilliance iCT; Product Code (REF): 728306; right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Philips North America Llc to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Philips North America Llc
Category
Health
Recall date
February 18, 2026
Units affected
Not reported
Sold at
Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX,
Recall number
FDA-Z-1315-2026
Made in
United States
Issuing agency
FDA

In context

  • Philips North America Llc has 34 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 356th Health recall we've logged in 2026.
Read the official FDA notice