MAQUET GMBH 2025-09-03 Health serious FDA
Health · FDA · Recall #FDA-Z-2420-2025

Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4)… Recall

Issued September 3, 2025 · MAQUET GMBH

In September 2025, MAQUET GMBH recalled the Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4)…, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.

What you should do

  1. 1 Stop using Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4)… right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact MAQUET GMBH to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
MAQUET GMBH
Category
Health
Recall date
September 3, 2025
Units affected
Not reported
Sold at
Worldwide - US Nationwide distribution in the states of KY, NY, TX and the countries of Algeria, Australia, Austria, Bulgaria, Canada, Chin
Recall number
FDA-Z-2420-2025
Made in
Germany
Issuing agency
FDA

In context

  • MAQUET GMBH has 1 recall in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 691st Health recall we've logged in 2025.
Read the official FDA notice

Related recalls