Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4)… Recall
Issued September 3, 2025 · MAQUET GMBH
In September 2025, MAQUET GMBH recalled the Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4)…, a serious health recall reported by the FDA.
Compiled by Maya Ellison, RecallCheckr
What this means for you
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.
What you should do
- 1 Stop using Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4)… right away.
- 2 Set it aside where it won't be used by mistake.
- 3 Contact MAQUET GMBH to arrange your voluntary. See the official notice below for contact details.
- 4 Keep the product until your remedy is fully processed. You may need it.
Recall details
- Brand
- MAQUET GMBH
- Category
- Health
- Recall date
- September 3, 2025
- Units affected
- Not reported
- Sold at
- Worldwide - US Nationwide distribution in the states of KY, NY, TX and the countries of Algeria, Australia, Austria, Bulgaria, Canada, Chin
- Recall number
- FDA-Z-2420-2025
- Made in
- Germany
- Issuing agency
- FDA
In context
- MAQUET GMBH has 1 recall in our database.
- We're tracking 3,576 Health recalls in our database.
- This is the 691st Health recall we've logged in 2025.