RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Intuitive Surgical, Inc. 2022-11-02 Health serious FDA
Health · FDA · Recall #FDA-Z-0133-2023

da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon… Recall

Issued November 2, 2022 · Intuitive Surgical, Inc.

In November 2022, Intuitive Surgical, Inc. recalled the da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon…, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery

What you should do

  1. 1 Stop using da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon… right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Intuitive Surgical, Inc. to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Intuitive Surgical, Inc.
Category
Health
Recall date
November 2, 2022
Units affected
Not reported
Sold at
US: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD, TX, IA, MA, NE, AZ, WA, CO, AL, SC
Recall number
FDA-Z-0133-2023
Made in
United States
Issuing agency
FDA

In context

  • Intuitive Surgical, Inc. has 11 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 870th Health recall we've logged in 2022.
Read the official FDA notice

Related recalls