RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Beckman Coulter, Inc. 2023-05-24 Health serious FDA
Health · FDA · Recall #FDA-Z-1598-2023

DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic… Recall

Issued May 24, 2023 · Beckman Coulter, Inc.

In May 2023, Beckman Coulter, Inc. recalled the DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic…, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

What you should do

  1. 1 Stop using DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic… right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Beckman Coulter, Inc. to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Beckman Coulter, Inc.
Category
Health
Recall date
May 24, 2023
Units affected
Not reported
Sold at
Worldwide Distribution
Recall number
FDA-Z-1598-2023
Made in
United States
Issuing agency
FDA

In context

  • Beckman Coulter, Inc. has 8 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 903rd Health recall we've logged in 2023.
Read the official FDA notice

Related recalls