Waldemar Link GmbH & Co. KG (Mfg Site) 2026-03-18 Health serious FDA
Health · FDA · Recall #FDA-Z-1521-2026
Endo-Model Replacement Plateau; Item Number: 15-0027/11; Recall
Issued March 18, 2026 · Waldemar Link GmbH & Co. KG (Mfg Site)
In March 2026, Waldemar Link GmbH & Co. KG (Mfg Site) recalled the Endo-Model Replacement Plateau; Item Number: 15-0027/11;, a serious health recall reported by the FDA.
Compiled by Maya Ellison, RecallCheckr
What this means for you
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
What you should do
- 1 Stop using Endo-Model Replacement Plateau; Item Number: 15-0027/11; right away.
- 2 Set it aside where it won't be used by mistake.
- 3 Contact Waldemar Link GmbH & Co. KG (Mfg Site) to arrange your voluntary. See the official notice below for contact details.
- 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):
Recall details
- Brand
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Category
- Health
- Recall date
- March 18, 2026
- Units affected
- Not reported
- Sold at
- Worldwide - US Nationwide distribution in the state of NJ and the countries of Belarus, Canada, Switzerland , Colombia, Germany, Spain, Uni
- Recall number
- FDA-Z-1521-2026
- Made in
- Germany
- Issuing agency
- FDA
In context
- Waldemar Link GmbH & Co. KG (Mfg Site) has 8 recalls in our database.
- We're tracking 3,576 Health recalls in our database.
- This is the 356th Health recall we've logged in 2026.