RecallCheckr

Critical recall: Death or serious injury has been reported or is considered likely. Stop using this product now.

DATEX--OHMEDA, INC. 2024-10-23 Health critical FDA
Health · FDA · Recall #FDA-Z-0047-2025

GE Heathcare Giraffe OmniBed; neo natal incubator Recall

Issued October 23, 2024 · DATEX--OHMEDA, INC.

In October 2024, DATEX--OHMEDA, INC. recalled the GE Heathcare Giraffe OmniBed; neo natal incubator, a critical health recall reported by the FDA. Affected owners are eligible for a voluntary. Here is what the hazard is and exactly what to do about it.

Compiled by Maya Ellison, RecallCheckr

What this means for you

GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.

What you should do

  1. 1 Stop using GE Heathcare Giraffe OmniBed; neo natal incubator right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact DATEX--OHMEDA, INC. to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
DATEX--OHMEDA, INC.
Category
Health
Recall date
October 23, 2024
Units affected
Not reported
Sold at
Worldwide distribution.
Recall number
FDA-Z-0047-2025
Made in
United States
Issuing agency
FDA

In context

  • DATEX--OHMEDA, INC. has 2 recalls in our database, including 2 rated Critical.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 737th Health recall we've logged in 2024.
Read the official FDA notice

Related recalls