RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

DePuy Orthopaedics, Inc. 2022-01-19 Health serious FDA
Health · FDA · Recall #FDA-Z-0487-2022

GLOBAL UNITE REV STEM SZ 10 Product Code: 110010600 Recall

Issued January 19, 2022 · DePuy Orthopaedics, Inc.

In January 2022, DePuy Orthopaedics, Inc. recalled the GLOBAL UNITE REV STEM SZ 10 Product Code: 110010600, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage

What you should do

  1. 1 Stop using GLOBAL UNITE REV STEM SZ 10 Product Code: 110010600 right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact DePuy Orthopaedics, Inc. to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
DePuy Orthopaedics, Inc.
Category
Health
Recall date
January 19, 2022
Units affected
Not reported
Sold at
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Denmark, Estonia, Finland, France, Germany
Recall number
FDA-Z-0487-2022
Made in
United States
Issuing agency
FDA

In context

  • DePuy Orthopaedics, Inc. has 4 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 870th Health recall we've logged in 2022.
Read the official FDA notice

Related recalls