Human Care USA, Inc. 2022-04-06 Health serious FDA
Health · FDA · Recall #FDA-Z-0852-2022
Hanger-Bar 2P 45cm Recall
Issued April 6, 2022 · Human Care USA, Inc.
In April 2022, Human Care USA, Inc. recalled the Hanger-Bar 2P 45cm, a serious health recall reported by the FDA.
Compiled by Maya Ellison, RecallCheckr
What this means for you
The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.
What you should do
- 1 Stop using Hanger-Bar 2P 45cm right away.
- 2 Set it aside where it won't be used by mistake.
- 3 Contact Human Care USA, Inc. to arrange your voluntary. See the official notice below for contact details.
- 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):
Recall details
- Brand
- Human Care USA, Inc.
- Category
- Health
- Recall date
- April 6, 2022
- Units affected
- Not reported
- Sold at
- U.S. Nationwide distribution in the states of CA, IL, LA, MA, NV, OH, OK, PA, TX, UT, VA, and WV.
- Recall number
- FDA-Z-0852-2022
- Made in
- United States
- Issuing agency
- FDA
In context
- Human Care USA, Inc. has 1 recall in our database.
- We're tracking 3,576 Health recalls in our database.
- This is the 870th Health recall we've logged in 2022.