Infusomat 60DROP METRISET PUMP SET, 3 SAFELINE-Used with an electrically-powered… Recall
Issued July 24, 2024 · B. Braun Medical, Inc.
In July 2024, B. Braun Medical, Inc. recalled the Infusomat 60DROP METRISET PUMP SET, 3 SAFELINE-Used with an electrically-powered…, a serious health recall reported by the FDA.
Compiled by Maya Ellison, RecallCheckr
What this means for you
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
What you should do
- 1 Stop using Infusomat 60DROP METRISET PUMP SET, 3 SAFELINE-Used with an electrically-powered… right away.
- 2 Set it aside where it won't be used by mistake.
- 3 Contact B. Braun Medical, Inc. to arrange your voluntary. See the official notice below for contact details.
- 4 Keep the product until your remedy is fully processed. You may need it.
Recall details
- Brand
- B. Braun Medical, Inc.
- Category
- Health
- Recall date
- July 24, 2024
- Units affected
- Not reported
- Sold at
- Worldwide distribution - US Nationwide and the country of Canada.
- Recall number
- FDA-Z-2380-2024
- Made in
- United States
- Issuing agency
- FDA
In context
- B. Braun Medical, Inc. has 8 recalls in our database.
- We're tracking 3,576 Health recalls in our database.
- This is the 737th Health recall we've logged in 2024.