RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Philips North America Llc 2024-05-29 Health serious FDA
Health · FDA · Recall #FDA-Z-1851-2024

Ingenia Ambition X- Magnetic Resonance (MR) systems are Medical Electrical Systems… Recall

Issued May 29, 2024 · Philips North America Llc

In May 2024, Philips North America Llc recalled the Ingenia Ambition X- Magnetic Resonance (MR) systems are Medical Electrical Systems…, a serious health recall reported by the FDA. Affected owners are eligible for a voluntary. Here is what the hazard is and exactly what to do about it.

Compiled by Maya Ellison, RecallCheckr

What this means for you

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

What you should do

  1. 1 Stop using Ingenia Ambition X- Magnetic Resonance (MR) systems are Medical Electrical Systems… right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Philips North America Llc to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Philips North America Llc
Category
Health
Recall date
May 29, 2024
Units affected
Not reported
Sold at
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, Fra
Recall number
FDA-Z-1851-2024
Made in
United States
Issuing agency
FDA

In context

  • Philips North America Llc has 34 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 737th Health recall we've logged in 2024.
Read the official FDA notice