RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Landauer 2024-01-03 Health serious FDA
Health · FDA · Recall #FDA-Z-0616-2024

Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: … Recall

Issued January 3, 2024 · Landauer

In January 2024, Landauer recalled the Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: …, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.

What you should do

  1. 1 Stop using Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: … right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Landauer to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Landauer
Category
Health
Recall date
January 3, 2024
Units affected
Not reported
Sold at
US and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam
Recall number
FDA-Z-0616-2024
Made in
United States
Issuing agency
FDA

In context

  • Landauer has 10 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 737th Health recall we've logged in 2024.
Read the official FDA notice

Related recalls