Micro-X Ltd. 2024-12-18 Health serious FDA
Health · FDA · Recall #FDA-Z-0664-2025
MICRO-X Rover Mobile X-ray System, # MXU-RV35 Recall
Issued December 18, 2024 · Micro-X Ltd.
In December 2024, Micro-X Ltd. recalled the MICRO-X Rover Mobile X-ray System, # MXU-RV35, a serious health recall reported by the FDA.
Compiled by Maya Ellison, RecallCheckr
What this means for you
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
What you should do
- 1 Stop using MICRO-X Rover Mobile X-ray System, # MXU-RV35 right away.
- 2 Set it aside where it won't be used by mistake.
- 3 Contact Micro-X Ltd. to arrange your voluntary. See the official notice below for contact details.
- 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):
Recall details
- Brand
- Micro-X Ltd.
- Category
- Health
- Recall date
- December 18, 2024
- Units affected
- Not reported
- Sold at
- Worldwide and US Nationwide Distribution: NJ, PR, CA
- Recall number
- FDA-Z-0664-2025
- Made in
- Australia
- Issuing agency
- FDA
In context
- Micro-X Ltd. has 3 recalls in our database.
- We're tracking 3,576 Health recalls in our database.
- This is the 737th Health recall we've logged in 2024.