RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Micro-X Ltd. 2024-12-18 Health serious FDA
Health · FDA · Recall #FDA-Z-0664-2025

MICRO-X Rover Mobile X-ray System, # MXU-RV35 Recall

Issued December 18, 2024 · Micro-X Ltd.

In December 2024, Micro-X Ltd. recalled the MICRO-X Rover Mobile X-ray System, # MXU-RV35, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

What you should do

  1. 1 Stop using MICRO-X Rover Mobile X-ray System, # MXU-RV35 right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Micro-X Ltd. to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Micro-X Ltd.
Category
Health
Recall date
December 18, 2024
Units affected
Not reported
Sold at
Worldwide and US Nationwide Distribution: NJ, PR, CA
Recall number
FDA-Z-0664-2025
Made in
Australia
Issuing agency
FDA

In context

  • Micro-X Ltd. has 3 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 737th Health recall we've logged in 2024.
Read the official FDA notice