RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Shimadzu Medical Systems 2022-04-13 Health serious FDA
Health · FDA · Recall #FDA-Z-0869-2022

MODEL: X-RAY TV SYSSTEM SONIALVISION safire17 Recall

Issued April 13, 2022 · Shimadzu Medical Systems

In April 2022, Shimadzu Medical Systems recalled the MODEL: X-RAY TV SYSSTEM SONIALVISION safire17, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.

What you should do

  1. 1 Stop using MODEL: X-RAY TV SYSSTEM SONIALVISION safire17 right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Shimadzu Medical Systems to arrange your fda mandated. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (FDA Mandated):

Recall details

Brand
Shimadzu Medical Systems
Category
Health
Recall date
April 13, 2022
Units affected
Not reported
Sold at
U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN
Recall number
FDA-Z-0869-2022
Made in
United States
Issuing agency
FDA

In context