RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Landauer 2023-09-20 Health serious FDA
Health · FDA · Recall #FDA-Z-2553-2023

nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring… Recall

Issued September 20, 2023 · Landauer

In September 2023, Landauer recalled the nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring…, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

What you should do

  1. 1 Stop using nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring… right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Landauer to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Landauer
Category
Health
Recall date
September 20, 2023
Units affected
Not reported
Sold at
Worldwide
Recall number
FDA-Z-2553-2023
Made in
United States
Issuing agency
FDA

In context

  • Landauer has 10 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 903rd Health recall we've logged in 2023.
Read the official FDA notice

Related recalls