Zimmer, Inc. 2026-07-01 Health serious FDA
Health · FDA · Recall #FDA-Z-2528-2026
Off-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide … Recall
Issued July 1, 2026 · Zimmer, Inc.
In July 2026, Zimmer, Inc. recalled the Off-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide …, a serious health recall reported by the FDA.
Compiled by Dana Whitfield, RecallCheckr
What this means for you
Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
What you should do
- 1 Stop using Off-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide … right away.
- 2 Set it aside where it won't be used by mistake.
- 3 Contact Zimmer, Inc. to arrange your voluntary. See the official notice below for contact details.
- 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):
Recall details
- Brand
- Zimmer, Inc.
- Category
- Health
- Recall date
- July 1, 2026
- Units affected
- Not reported
- Sold at
- US Nationwide distribution.
- Recall number
- FDA-Z-2528-2026
- Made in
- United States
- Issuing agency
- FDA
In context
- Zimmer, Inc. has 6 recalls in our database.
- We're tracking 3,673 Health recalls in our database.
- This is the 453rd Health recall we've logged in 2026.