RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Microbiologics Inc 2022-02-16 Health serious FDA
Health · FDA · Recall #FDA-Z-0587-2022

QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains… Recall

Issued February 16, 2022 · Microbiologics Inc

In February 2022, Microbiologics Inc recalled the QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains…, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.

What you should do

  1. 1 Stop using QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains… right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Microbiologics Inc to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Microbiologics Inc
Category
Health
Recall date
February 16, 2022
Units affected
Not reported
Sold at
Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia,
Recall number
FDA-Z-0587-2022
Made in
United States
Issuing agency
FDA

In context

  • Microbiologics Inc has 7 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 870th Health recall we've logged in 2022.