What this means for you

Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.

What you should do

  1. 1 Stop using RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097. right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Siemens Healthcare Diagnostics Inc to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Siemens Healthcare Diagnostics Inc
Category
Health
Recall date
August 6, 2025
Units affected
Not reported
Sold at
worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.
Recall number
FDA-Z-2230-2025
Made in
United States
Issuing agency
FDA

In context

  • Siemens Healthcare Diagnostics Inc has 2 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 691st Health recall we've logged in 2025.
Read the official FDA notice

Related recalls