SIGNA Premier systems Recall
Issued April 8, 2026 · GE Healthcare LLC
In April 2026, GE Healthcare LLC recalled the SIGNA Premier systems, a serious health recall reported by the FDA.
Compiled by Maya Ellison, RecallCheckr
What this means for you
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
What you should do
- 1 Stop using SIGNA Premier systems right away.
- 2 Set it aside where it won't be used by mistake.
- 3 Contact GE Healthcare LLC to arrange your voluntary. See the official notice below for contact details.
- 4 Keep the product until your remedy is fully processed. You may need it.
Recall details
- Brand
- GE Healthcare LLC
- Category
- Health
- Recall date
- April 8, 2026
- Units affected
- Not reported
- Sold at
- Worldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United
- Recall number
- FDA-Z-1767-2026
- Made in
- United States
- Issuing agency
- FDA
In context
- GE Healthcare LLC has 1 recall in our database.
- We're tracking 3,576 Health recalls in our database.
- This is the 356th Health recall we've logged in 2026.