RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

GE Healthcare LLC 2026-04-08 Health serious FDA
Health · FDA · Recall #FDA-Z-1767-2026

SIGNA Premier systems Recall

Issued April 8, 2026 · GE Healthcare LLC

In April 2026, GE Healthcare LLC recalled the SIGNA Premier systems, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.

What you should do

  1. 1 Stop using SIGNA Premier systems right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact GE Healthcare LLC to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
GE Healthcare LLC
Category
Health
Recall date
April 8, 2026
Units affected
Not reported
Sold at
Worldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United
Recall number
FDA-Z-1767-2026
Made in
United States
Issuing agency
FDA

In context

  • GE Healthcare LLC has 1 recall in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 356th Health recall we've logged in 2026.
Read the official FDA notice