RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Medtronic Navigation, Inc. 2024-10-16 Health serious FDA
Health · FDA · Recall #FDA-Z-0050-2025

StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation… Recall

Issued October 16, 2024 · Medtronic Navigation, Inc.

In October 2024, Medtronic Navigation, Inc. recalled the StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation…, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal.

What you should do

  1. 1 Stop using StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation… right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Medtronic Navigation, Inc. to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Medtronic Navigation, Inc.
Category
Health
Recall date
October 16, 2024
Units affected
Not reported
Sold at
Worldwide - U.S Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, M
Recall number
FDA-Z-0050-2025
Made in
United States
Issuing agency
FDA

In context

  • Medtronic Navigation, Inc. has 1 recall in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 737th Health recall we've logged in 2024.
Read the official FDA notice

Related recalls