In November 2023, Aizu Olympus Co., Ltd. recalled the Tracheal fiberscope-airway management, which includes observation to access airway…, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

What you should do

  1. 1 Stop using Tracheal fiberscope-airway management, which includes observation to access airway… right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Aizu Olympus Co., Ltd. to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Aizu Olympus Co., Ltd.
Category
Health
Recall date
November 29, 2023
Units affected
Not reported
Sold at
Worldwide - US Nationwide distribution.
Recall number
FDA-Z-0338-2024
Made in
Japan
Issuing agency
FDA

In context

  • Aizu Olympus Co., Ltd. has 4 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 903rd Health recall we've logged in 2023.
Read the official FDA notice

Related recalls