RecallCheckr

Serious recall: Injury is possible or has been reported. Take action promptly.

Canon Medical System, USA, INC. 2026-01-21 Health serious FDA
Health · FDA · Recall #FDA-Z-1017-2026

VANTAGE GALAN 3T Model MRT-3020/MEXL-3020 Recall

Issued January 21, 2026 · Canon Medical System, USA, INC.

In January 2026, Canon Medical System, USA, INC. recalled the VANTAGE GALAN 3T Model MRT-3020/MEXL-3020, a serious health recall reported by the FDA.

Compiled by Maya Ellison, RecallCheckr

What this means for you

There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

What you should do

  1. 1 Stop using VANTAGE GALAN 3T Model MRT-3020/MEXL-3020 right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact Canon Medical System, USA, INC. to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
Canon Medical System, USA, INC.
Category
Health
Recall date
January 21, 2026
Units affected
Not reported
Sold at
U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
Recall number
FDA-Z-1017-2026
Made in
United States
Issuing agency
FDA

In context

  • Canon Medical System, USA, INC. has 2 recalls in our database.
  • We're tracking 3,576 Health recalls in our database.
  • This is the 356th Health recall we've logged in 2026.
Read the official FDA notice